The use of endobronchial valve device to eliminate air leak RESPIRATORY MEDICINE Fann, J. I., Berry, G. J., Burdon, T. A. 2006; 100 (8): 1402-1406

Abstract

We evaluated an endobronchial valve device in the treatment of surgically created air leak or pneumothorax by eliminating antegrade flow.Six sheep underwent general anesthesia with positive pressure ventilation and left thoracotomy. After division of the mediastinal pleura, the contralateral cranial lobe was identified and a 2.5 cmx1.5 cm laceration created with resultant air leak. Using bronchoscopy, we deployed a valve device in the bronchus of the injured segment. Chest drainage tube was placed and the thoracotomy closed. At 1 week (n=3) and 4 weeks (n=3), the animals underwent general anesthesia, bronchoscopy and right thoracotomy.All animals survived the procedure. Bronchoscopic valve device placement in the segmental bronchus resolved the air leak immediately. After closure of thoracotomy, the chest tube demonstrated minimal drainage with no air leak. At 1 and 4 weeks, bronchoscopy showed no change in device location, and the treated segments were atelectatic with fibrous scar at the injured site.Collapse of a selected lung segment with resolution of air leak can be achieved using bronchoscopically implanted valve device. The valve device may facilitate treatment of patients with post-surgical or post-traumatic persistent air leak.

View details for DOI 10.1016/j.rmed.2005.11.011.