Invest Radiol. 1994 Dec;29(12):1038-42.

Evaluation of myocardial motion tracking with cine-phase contrast magnetic resonance imaging.

Author information

1
Department of Radiology, Stanford University School of Medicine, California, USA.

Abstract

RATIONALE AND OBJECTIVES:

The accuracy of myocardial motion measurements, computed from cine-phase contrast (cine-PC) magnetic resonance (MR) velocity data, was compared with directly visualized motion of MR signal voids caused by implanted tantalum markers in anesthetized dogs.

METHODS:

Magnetic resonance imaging (MRI) data were electrocardiogram-gated and divided into 16 phases per cardiac cycle. Myocardial trajectories as a function of time in the cardiac cycle were measured using both methods for four to seven markers in each of eight animals.

RESULTS:

The peak observed in-plane excursion was 4.0 +/- 2.1 mm. The average deviation between displacements derived from velocity data versus displacements visualized directly was 1.1 +/- 0.7 mm (27.5% of the peak displacement). The difference was less if three separate MR scans were used to measure each velocity component in the cine-PC method. This improvement is probably caused by improved temporal resolution.

CONCLUSIONS:

Cine-PC MRI offers a noninvasive method for accurate quantification of myocardial motion.

PMID:
7721545