Format

Send to

Choose Destination
Magn Reson Med. 1998 Apr;39(4):581-7.

Gradient characterization using a Fourier-transform technique.

Author information

1
Department of Radiology, Stanford University, California 94305-5488, USA.

Abstract

This paper describes a technique for characterizing the gradient subsystem of a magnetic resonance (MR) system. The technique uses a Fourier-transform analysis to directly measure the k-space trajectory produced by an arbitrary gradient waveform. In addition, the method can be easily extended to multiple dimensions and can be adapted to measuring residual gradient effects such as eddy currents. Several examples of gradient waveform and eddy-current measurements are presented. Also, it is demonstrated how the eddy-current measurements can be parameterized with an impulse-response formalism for later use in system tuning. When compared to a peak-fitting analysis, this technique provides a more direct extraction of the k-space measurements, which reduces the possibility of analysis error. This approach also has several advantages as compared to the conventional eddy-current measurement technique, including the ability to measure very short time constant effects.

PMID:
9543420
[Indexed for MEDLINE]

Supplemental Content

Full text links

Icon for Wiley
Loading ...
Support Center