Beam line 4-3 is a wiggler side-station dedicated for X-ray absorption spectroscopy and EXAFS measurements on biological, environmental, catalysis, and materials systems. This station enables tender X-ray measurement (S K-edge and up) and is setup with a He flight path from the beam line optics to the sample. The beam is collimated and unfocused to allow for high energy-resolution measurements on homogenous samples. BL4-3 is equipped with a 4-element solid-state silicon drift detector in addition to ionization chambers and PIPS/Lytle detectors. A LHe cryostream allows for low-temperature measurements. A motorized sample holder stage allows sequential measurement on multiple samples.
Status | Open |
||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|
Supported Techniques | X-ray Absorption Spectroscopy |
||||||||||
Main Scientific Disciplines | Bioscience Environmental Sciences Materials Sciences |
||||||||||
Beam Line Specifications | |||||||||||
Source20-pole, 2.0-Tesla Wiggler, 0.75 mrad, Side station
OpticsM0 mirror: vertically collimating, flat, bent, 1.0 m, Si, Ni-coated, cutoff 4.5 to 14.5 keV The Ni coating on M0 prohibits Ni K- and in certain cases Cu K-edge XAS measurement. A Cr underlayer in M0 mirror restricts applications at the Cr K-edge. Contact responsible beamline staff for more information. Monochromator: LN2 cooled, Si(111), ϕ = 0o & 90o. double-crystal, non-fixed exit slit Monochromator Crystal Glitch Library Crystal changes need to be scheduled and coordinated in advance with BL support staff Absorption127 micron Be |
|||||||||||
Instrumentation | Detectors: Ionization chambers, PIPS, Lytle, Hitachi HTA 4-element solid-state Si drift detector, electron-yield detector Cryostat: LHe cryostream Other: He-purged sample box. He flow meters and oxygen content meter on sample space. Automated shutter to alleviate beam damage. Motorized multiple solid-sample holder |
||||||||||
Sample Environment | Ambient He atmosphere in gas tight sample box (room temperature) plastic sample compartment (for He cryostream) Sample temperature: 20-50 K (He cryostream) or Room temperature (no cryostat) Sample size: 5 X 20 mm Minimum Sample Requirement: 2-5 mM, 200 ul or 100 ppm, 2-5 mg at tender energies. 2 mM, 200 μl solutions or 10 ppm, 10 mg solids at higher energies |
||||||||||
Data Acquisition and Analysis | XASCollect, Xasscan, EXAFSPAK, SixPAK |
||||||||||
Beam Line Phone | 650-926-5243 . On-site Users: Contact the Duty Operator at 9-926-4040 |
||||||||||
Beam Line Contact | |||||||||||
Beam Line Engineering Notes |